As a compact, pressure-resistant pH probe in analytical laboratories or as one of many measuring channels in SMB systems or in process chromatography.
Accurate, temperature-compensated measurement of the pH value. Two measuring cells with different bores to choose from.
Sensors | pH measure | ISFET |
---|---|---|
pH reference | KCl or Ag/AgCl | |
temperature | Pt-1000 | |
pH | display range (1) | 0 ... 14 (2) |
linear range | 2 ... 12 | |
resolution | 0.001 | |
accuracy | ±0.1 | |
sampling rate | 1 ... 10 Hz | |
Temperature | display range | 0 ... 100 °C |
resolution | 0.1 °C | |
accuracy | ±0.2 °C | |
Liquid path | bore | ID = 1.0 mm |
dwell volume | 60 µl | |
flow rate | up to 1,000 ml/min | |
back pressure | 1 bar at 500 ml/min (3) |
Process | temperature | 5 ... 80 °C |
---|---|---|
pressure | 20 bar (290 psi) | |
Liquid path | ports | 1/4″-28 UNF flat-bottom |
Materials | cell body | PEEK (4) |
wetted | PEEK (4), SiO2, FKM | |
Interface | USB-C or RS-485 (data, energy) | |
Power | electrical | < 2,5 W ( < 500 mA at 5 V) |
Protocol | open, text-based | |
Drivers | DataApex Clarity, SCPA PrepCon, LabView (5) | |
Markings | CE, UKCA |
Sensors | pH measure | ISFET |
---|---|---|
pH reference | KCl or Ag/AgCl | |
Temperature | Pt-1000 | |
pH | display range (1) | 0 ... 14 (2) |
linear range | 2 ... 12 | |
resolution | 0.001 | |
accuracy | ±0.1 | |
sampling rate | 1 ... 10 Hz | |
Temperature | display range | 0 ... 100 °C |
resolution | 0.1 °C | |
accuracy | ±0.2 °C | |
Liquid path | bore | ID = 2.2 mm |
dwell volume | 100 µl | |
flow rate | up to 1,000 ml/min | |
back pressure | < 1 bar at 500 ml/min (3) |
Process | temperature | 5 ... 80 °C |
---|---|---|
pressure | 20 bar (290 psi) | |
Liquid path | ports | 1/4″-28 UNF flat-bottom |
Materials | cell body | PEEK (4) |
wetted | PEEK (4), SiO2, FKM | |
Interface | USB-C or RS-485 (data, energy) | |
Power | electrical | < 2,5 W ( < 500 mA at 5 V) |
Protocol | open, text-based | |
Drivers | DataApex Clarity, SCPA PrepCon, LabView (5) | |
Markings | CE, UKCA |